Welcome To TAKTVOLL

ES-200PK Electrosurgical Generator for vet

Short Description:

ES-200PK is a Multifunctional electrosurgical generator with a wide range of application departments and a very high-cost performance. It uses a new generation of tissue density instant feedback technology, which can automatically adjust the output power according to the change in tissue density. Especially suitable for veterinary medicine use.


Product Detail

Product Tags

Feature

3 monopolar cutting modes: pure cut, blend 1, blend 2
Pure cut: cut the tissue cleanly and accurately without coagulation
blend 1: Use when the cutting speed is slightly slow and a small amount of hemostasis is required.
blend 2: Compared with blend 1, it is used when the cutting speed is slightly slower and the better hemostatic effect is needed.

3 monopolar coagulation modes: spray coagulation, forced coagulation, and soft coagulation
spray coagulation: high-efficiency coagulation without contact surface. The coagulation depth is shallow. The tissue is removed by evaporation. It usually uses a Blade or ball electrode for coagulation.
forced coagulation: It’s non-contact coagulation. The output threshold voltage is lower than spray coagulation. It’s suitable for coagulation in a small area.
soft coagulation: Mild coagulation penetrates deeply to prevent tissue carbonization and reduce electrode adhesion to tissue.

2 bipolar output modes: standard and fine
Standard mode: It’s suitable for most bipolar applications. Keep the low voltage to prevent sparks.
Fine mode: It’s used for high precision and fine control of drying amount. Keep the low voltage to prevent sparks.

CQM contact quality monitoring system
Automatically monitor the quality of contact between the dispersive pad and the patient in real-time. If the contact quality is lower than the set value, there will be a sound and light alarm and cut off the power output to ensure safety.

Electrosurgical pens and foot switch control

Start with the recently used mode, power, and other parameters
Volume adjustment function.

Cut and coagulate in an intermittent manner.

Functional self-test
After each turn-on, the high-frequency electrosurgical unit will immediately execute a self-test procedure. Once the internal abnormality of the system is found and the self-test fails, the current output will be automatically cut off immediately. This ensures that the ES-200PK generator is always in good working condition and performance. During the self-test, it is also tested whether the connected accessories are functioning normally.